Pavelka, R.

Link to this page

Authority KeyName Variants
9b174d42-04ee-45bb-9e63-3f759f1a23ee
  • Pavelka, R. (1)
Projects

Author's Bibliography

Measurement of the human cadaver ossicle vibration amplitude by fiber-optic interferometry

Đinović, Zoran; Tomić, Miloš; Pavelka, R.; Sprinzl, G.; Traxler, H.

(Rijeka : Croatian Society for Information, Communication and Electronic Technology - MIPRO, 2020)

TY  - CONF
AU  - Đinović, Zoran
AU  - Tomić, Miloš
AU  - Pavelka, R.
AU  - Sprinzl, G.
AU  - Traxler, H.
PY  - 2020
UR  - https://dais.sanu.ac.rs/123456789/9985
AB  - In this paper we present the results obtained by contactless measurement of vibration of the incus inside the middle ear of a human cadaver, exposed to an external calibrated sound source. The measurement configuration was a Michelson fiber-optic interferometer based on a 3x3 fiber optic coupler, with a VCSEL source. The interferometric signal fading was solved using two quasiquadrature signals and specialized signal processing. The ossicle reflection was augmented by a very light piece of retroreflective material attached to the bone. We measured the amplitude of vibration at different sound pressure levels, independently monitored by a decibel meter placed in the proximity of the cadaver ear. The accuracy of the measurement technique was proved by using a calibrated piezo transducer exposed to vibrate with amplitude of 7nm at 1kHz. We performed the experiments on 8 cadavers in total. The measured amplitudes were found to be in the range between 1 pm and 1 nm, under the sound excitation of 40-90 dB SPL @ 1kHz, respectively.
PB  - Rijeka : Croatian Society for Information, Communication and Electronic Technology - MIPRO
C3  - Proceedings / MIPRO 2020 : 43rd International Convention, September 28, 2020 - October 2, 2020, Opatija, Croatia
T1  - Measurement of the human cadaver ossicle vibration amplitude by fiber-optic interferometry
SP  - 2232
EP  - 2236
UR  - https://hdl.handle.net/21.15107/rcub_dais_9985
ER  - 
@conference{
author = "Đinović, Zoran and Tomić, Miloš and Pavelka, R. and Sprinzl, G. and Traxler, H.",
year = "2020",
abstract = "In this paper we present the results obtained by contactless measurement of vibration of the incus inside the middle ear of a human cadaver, exposed to an external calibrated sound source. The measurement configuration was a Michelson fiber-optic interferometer based on a 3x3 fiber optic coupler, with a VCSEL source. The interferometric signal fading was solved using two quasiquadrature signals and specialized signal processing. The ossicle reflection was augmented by a very light piece of retroreflective material attached to the bone. We measured the amplitude of vibration at different sound pressure levels, independently monitored by a decibel meter placed in the proximity of the cadaver ear. The accuracy of the measurement technique was proved by using a calibrated piezo transducer exposed to vibrate with amplitude of 7nm at 1kHz. We performed the experiments on 8 cadavers in total. The measured amplitudes were found to be in the range between 1 pm and 1 nm, under the sound excitation of 40-90 dB SPL @ 1kHz, respectively.",
publisher = "Rijeka : Croatian Society for Information, Communication and Electronic Technology - MIPRO",
journal = "Proceedings / MIPRO 2020 : 43rd International Convention, September 28, 2020 - October 2, 2020, Opatija, Croatia",
title = "Measurement of the human cadaver ossicle vibration amplitude by fiber-optic interferometry",
pages = "2232-2236",
url = "https://hdl.handle.net/21.15107/rcub_dais_9985"
}
Đinović, Z., Tomić, M., Pavelka, R., Sprinzl, G.,& Traxler, H.. (2020). Measurement of the human cadaver ossicle vibration amplitude by fiber-optic interferometry. in Proceedings / MIPRO 2020 : 43rd International Convention, September 28, 2020 - October 2, 2020, Opatija, Croatia
Rijeka : Croatian Society for Information, Communication and Electronic Technology - MIPRO., 2232-2236.
https://hdl.handle.net/21.15107/rcub_dais_9985
Đinović Z, Tomić M, Pavelka R, Sprinzl G, Traxler H. Measurement of the human cadaver ossicle vibration amplitude by fiber-optic interferometry. in Proceedings / MIPRO 2020 : 43rd International Convention, September 28, 2020 - October 2, 2020, Opatija, Croatia. 2020;:2232-2236.
https://hdl.handle.net/21.15107/rcub_dais_9985 .
Đinović, Zoran, Tomić, Miloš, Pavelka, R., Sprinzl, G., Traxler, H., "Measurement of the human cadaver ossicle vibration amplitude by fiber-optic interferometry" in Proceedings / MIPRO 2020 : 43rd International Convention, September 28, 2020 - October 2, 2020, Opatija, Croatia (2020):2232-2236,
https://hdl.handle.net/21.15107/rcub_dais_9985 .